1 documents found
Information × Registration Number 0214U006460, 0108U002447 , R & D reports Title Development of precise humidity sensor for natural gas analysis. popup.stage_title Head Venger Evgeniy Fedorovich, Registration Date 17-04-2014 Organization Institute of Semiconductor Physics of National Academy of Sciences of Ukrain popup.description2 1. Developed and made model of block of electronics for a hygrometer with a hand management in composition the microprocessor block of management, electronic module of display and module of electric feed of hygrometer. His experimental researches are conducted. For communication of measuring data from a hygrometer to the computer developed and made model of interface electric board and flow line. A flow line is spark-proof and can be utillized in explosive environments, such as the gas stations. 2. The analysis of basic physical processes which flow in the condensation hygrometers and influence of these processes on exactness of measuring dew point is conducted. The new method of measuring of point of dew of gases is offered. In a new method influence of speed of cooling of mirror is taken into account on exactness of measuring. As compared to analogues a method allows substantially to promote exactness of measuring of dew point on moisture. Product Description popup.authors Шапар Володимир Миколайович popup.nrat_date 2020-04-02 Close
R & D report
Head: Venger Evgeniy Fedorovich. Development of precise humidity sensor for natural gas analysis.. (popup.stage: ). Institute of Semiconductor Physics of National Academy of Sciences of Ukrain. № 0214U006460
1 documents found