Information × Registration Number 0214U006486, 0113U003680 , R & D reports Title Development of stationary and portable samples of sensors for low-energy X-ray and gamma- radiation detecting. popup.stage_title Head Brodyn Mykhajlo Semenovych, Registration Date 30-04-2014 Organization Institute of physics NASU popup.description2 Designed and manufactured of the laboratory layouts radiometric detector of ionizing radiation consists of the unit electronics and the measurement unit. The device can be used for measuring the intensity of the weak flow of ionizing radiation. Novelty design - use as sensors for ionizing radiation semiconductor zinc selenide (ZnSe) single crystals using direct conversion of gamma- quantum energy into an electrical signal. Consequently, an efficiently working logger environment at temperatures up to 150 0C. Conducted testing of the device in pulsed operation mode, using a source of radiation radionuclide Ra226: the outputs of both detectors obtained clearly shaped pulses The detector can be used in scientific, medical and environmental studies for radiometric and dosimetric control in industrial plants, radiation safety inspection of premises and buildings.5481 Product Description popup.authors Бродин Михайло Семенович Будник Петро Іванович Весна Валерій Терентійович Дегода Володимир Якович Зайцевський Ігор Лаврович Кожушко Богдан Васильович Шевченко Сергій Віталійович popup.nrat_date 2020-04-02 Close
R & D report
Head: Brodyn Mykhajlo Semenovych. Development of stationary and portable samples of sensors for low-energy X-ray and gamma- radiation detecting.. (popup.stage: ). Institute of physics NASU. № 0214U006486