1 documents found
Information × Registration Number 0209U001628, 0108U003552 , R & D reports Title Reconstruction of scintillation detector signal for application in PET (positron-emissive tomograph) popup.stage_title Head Pedash A,Y., Registration Date 18-02-2009 Organization Scientific-research department of alkali halide crystals with experimental manufacture STC "Institute for single crystals" NAS of Ukraine popup.description2 With help of computer simulation method positional sensitivity of uniform annular scintillation detector was researched. Research of three variants for detector construction with different shape of scintillation crystal and geometry of photodetectors arrangement was carried out. It was shown that annular detector with square section has the best angular sensitivity to the location of scintillation point. Sensitivity to displacement of interaction point in radial direction and to the side of end faces in this configuration was insignificant.5635 Product Description popup.authors popup.nrat_date 2020-04-02 Close
R & D report
Head: Pedash A,Y.. Reconstruction of scintillation detector signal for application in PET (positron-emissive tomograph). (popup.stage: ). Scientific-research department of alkali halide crystals with experimental manufacture STC "Institute for single crystals" NAS of Ukraine. № 0209U001628