1 documents found
Information × Registration Number 0214U002901, 0113U001827 , R & D reports Title Search for optimal physicotechnical solutions for creation of the radio-immune analyzer based on new scintillator - silicone PMT system. popup.stage_title Head Borodenko J.A., Registration Date 31-01-2014 Organization Scientific-research department of alkali halide crystals with experimental manufacture STC "Institute for single crystals" NAS of Ukraine popup.description2 For the first time for creation of a highly stable domestic radio-immune analyzer on the basis of scintillator - silicone PMT system in the detection block was applied a new type of photoconverter - silicon photomultiplier tube (Si- PMT). Detection block has been successfully tested, parameters that were laid down in the specifications were confirmed. А new system for detection of gamma - SC radiation - Si- PMT was developed. As a result, a layout of radio-immune analyzer was made. Recommendations were provided for creation of radio-immune analyzer. Product Description popup.authors Білогуб B. В. Селегенев Є. М. popup.nrat_date 2020-04-02 Close
R & D report
2
Head: Borodenko J.A.. Search for optimal physicotechnical solutions for creation of the radio-immune analyzer based on new scintillator - silicone PMT system.. (popup.stage: ). Scientific-research department of alkali halide crystals with experimental manufacture STC "Institute for single crystals" NAS of Ukraine. № 0214U002901
1 documents found